Page 1-62
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Master Alarm Power-Up
Chapter 1: Introduction
b. The second LED (CX) indicates whether the Ethernet is connected prop-
erly.
–
Solid ON indicates that the Ethernet is wired properly.
–
Solid OFF indicates that the wiring is bad or that the network is not
activated.
c. The third LED (CY) indicates when an Ethernet packet of information is
being sent or received. It flashes both when sending and receiving infor-
mation.
–
When the network is activated, sending and receiving information
starts.
5. If the Master Alarm is functioning properly, lock the front panel (DE) (see
figure 1-17 on page 1–60).
6. If the alarm has just been installed, or if any Area Alarm Modules communi-
cating with the Master Alarm have been removed or replaced, do the
Clear
Network
function (refer to step 7 on page 1-67).
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...