MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Configuring the Master Alarm Using the Front Panel
Chapter 1: Introduction
Table 1-29. Master Alarm Front Panel Configuration Mode
NOTE:
For each option, the display shows only one setting at a time.
1. When the display shows the first option,
Language
,
press the
Test key
to scroll through the following choices:
English
Francais
Espanol
2. Press the
Silence key
to select one.
3. When the display shows the next option,
Audible Return Time
,
press the
Test key
to scroll through the following choices:
Never
30 minutes
60 minutes
90 minutes
120 minutes
4. Press the
Silence key
to select one.
NOTE:
To comply with the requirements of the City of New York Department of
Buildings; Materials and Equipment Acceptance Division (M.E.A.), do not set
the alarm return time to “never.” Set the alarm to return every 30 minutes.
The next option is for the N2 Address. This is the Ethernet network address for
the optional connection to a building automation system.
Front Panel Item
Description
Test key
Scrolls through the settings for each option.
Silence key
Selects the settings for each option, and
then advances to the next option. Advance
more quickly by pressing and holding.
Display
Shows the options and settings.
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...