14. When the display shows the next option,
Input 1 is - System
,
press the
Test key
to scroll through the following choices:
1
2
3
4
5
6
7
8
9
15. Press the
Silence key
to select one.
NOTE:
Typically, all of the alarm inputs for the same type of gas or vacuum are mapped
to the same LED. The message on the display shows the specific source of the
alarm. When you are facing the front panel, the LEDs are numbered one through
nine.
1
2
3
4
5
6
7
8
9
16. When the display shows the next option,
Input 1 lights
LED
,
press the
Test key
to scroll through the following choices:
1
2
3
4
5
6
7
8
9
17. Press the
Silence key
to select one.
NOTE:
Only one type of gas can be assigned to an LED. After an LED has been
assigned to a type of gas, that LED number is available only for that particular
type of gas.
18. Repeat step 9 through step 17 in order. The display advances to the next
option, which is the
Gas Type
for the next consecutive input, until
Input 30
is configured.
NOTE:
After Input 30 is configured, the alarm returns to normal operation.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-69
Configuring the Master Alarm Using the Front Panel
Chapter 1: Introduction
1
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...