26
Part No. 6-847684-00 Rev. D01
Example 5: Area alarm panel - gas services for
oxygen, medical air, and vacuum.
I. C. U.
I. C. U.
I. C. U.
Oxygen
Sensor
Module
Connected to main
oxygen pipeline
Med. Air
Sensor
Module
Connected to main
med. air pipeline
Vacuum
Sensor
Module
Connected to main
vacuum pipeline
Oxygen Digital Display Module
Med. Air Digital Display Module
Vacuum Digital Display Module
pin 5 (+)
pin 6 (-)
pin 5 (+)
pin 6 (-)
pin 5 (+)
pin 6 (-)
Typical Master/Area Alarm Wiring Examples
NOTE:
Field wiring cable shields must be grounded at
only one end, inside alarm panel back box. Refer
to page 31 for details.