29
Part No. 6-847684-00 Rev. D01
1. Identify each pair of field installed
sensor wires inside alarm panel back
box assembly. Note polarity of each
wire from previously installed remote
sensor module.
2. Route each pair of sensor wires as
shown in Figure 18 to appropriate
digital display module. Each digital
display module is labeled with the gas
type. Verify the appropriate remote
sensor wires are connected to correct
digital display module.
Finish Installation
5. Tuck excess wiring into open space
behind sensor.
6. Mark location monitored by sensor in
space provided on sensor label. Refer
to Figure 17.
Local Sensors:
1. Remove sensor module from shipping
carton.
2. Insert sensor into valve opening in
pressure/vacuum service rough-in. An
indexing pin provides for proper
orientation and prevents gas service
cross-connection. Push sensor all the
way in and secure with (2) #6-32 x 1-
1/4” screws. Repeat this process for
all sensors within alarm panel.
3. The digital display module(s) located
on front panel assembly are factory
wired with a harness and polarized
two-pin connector for each sensor
module. Each digital display module
is labeled with the gas type. Match
gas type of each sensor to gas type of
each digital display module and join
polarized connectors of sensor and
digital display module harnesses.
4. Mark location monitored by sensor in
space provided on sensor label. Refer
to Figure 17.
Figure 17: Sensor Module Location Label
Remote Sensor Wiring
Figure 18: Remote Sensor Wire Routing
NOTE:
Remove the blue protective netting from the sen-
sor connector.
Location label
Holes provided for
entrance of remote wiring
Route wires behind
metal brace
Sensor Installation