31
Part No. 6-847684-00 Rev. D01
Figure 23: Multi-Signal Front Panel Labeling
Finish Installation
Source Equipment Wiring
Figure 21: Source Equipment Wire Routing
Holes provided for
entrance of remote wiring
Route wires behind
metal brace
Figure 22: Source Signal Wiring Termination
Multi-Signal Module
Signal 1
Signal 2
Signal 3
Signal 4
Signal 5
Sample label
(provided)
Slide label(s) into slits
in front panel overlay
Field Wiring Cable Shield Grounding
All field wiring cable shields must be
grounded inside alarm panel back box. A
grounding screw is provided in upper right
corner of alarm back box as shown in
Figure 24. Shields from several cables
may be wrapped together and crimped into
one ring lug (by others) .
Figure 24: Field Wiring Cable Shield Grounding
Cable shield
grounding screw
(crimp lug not provided)
CAUTION:
Keep shield wires to ground screw as short as
possible so they can not touch front panel circuit
boards when front panel is closed