37
Part No. 6-847684-00 Rev. D01
RR EE P
P TT
(audible alarm repeat)
Used to set time interval for
reactivation of audible alarm after
alarm MUTE
button has been
pressed. User choices are
OO FF FF
,
11
,
55
,
11 55
,
33 0
0
(minutes),
11 HH RR
,
8
8 HH RR
,
11 2
2 HH RR
or
2
2 44 HH RR
(hour). Factory setting is
OO FF FF
.
When set to
OO FF FF
, audible alarm will
not reactivate after an active alarm
has been silenced. Any new alarm will
reactivate audible alarm.
II NN TT SS
(display intensity)
Used to adjust digital module’s display
brightness. Factory setting is
maximum intensity.
Changing Communication Parameters -
(
SS EE TT 2
2
):
1. Press and HOLD SETUP button for
TWO seconds until
SS EE TT 2
2
is displayed.
2. Press and release TEST button on
front of alarm panel.
A
A D
D D
D RR
will be
displayed for one second and then
current network address. Select
desired address by pressing
!
or
"
buttons. Each module must have a
unique address.
3. Press and release TEST button.
BB A
A U
U D
D
will be displayed for one second
and then current baud rate. Select
desired baud rate by pressing
!
or
"
buttons.
4. Press and release TEST button.
NN D
D FF
will be displayed for one second and
Operation
Set-Up Procedure
NOTE:
If module is idle for more than one minute during
set-up procedure, module will chirp three times,
return to monitor mode and will default to previous
settings. No changes to settings will be saved.
Set-up must be completed and module must dis-
play
DD OO NN EE
in order for changes to be saved.
then current network data format.
Select desired data format by
pressing
!
or
"
buttons.
5. Press and release TEST button.
P
P D
D FF
will be displayed for one second and
then current pressure data format.
Select desired data format by
pressing
!
or
"
buttons.
6. Press and release TEST button to
save changes.
D
D OO NN EE
will be
displayed for two seconds. Digital
display module will return to
monitoring mode.
Changing Miscellaneous Parameters -
(
SS EE TT 33
):
1. Press and HOLD SETUP button for
FOUR seconds until
SS EE TT 33
is
displayed.
2. Press and release TEST button on
front of alarm panel.
SS EE RR U
U
will be
displayed for one second and then
current gas service identification.
Select desired gas service by
pressing
!
or
"
buttons.
3. Press and release TEST button.
OO FF FF SS
will be displayed for one second
and then current display offset. Select
offset by pressing
!
or
"
buttons.
4. Press and release TEST button.
RR EE P
P TT
will be displayed for one second
and then current audible alarm repeat
time. Select desired repeat time by
pressing
!
or
"
buttons.
5. Press and release TEST button.
II NN TT SS
will be displayed. Select desired
display brightness by pressing
!
or
"
buttons. Each push will change
brightness of display. Audible alarm
will chirp one time with each push and
two times when limit reached.
6. Press and release TEST button to
save changes.
D
D OO NN EE
will be
displayed for two seconds. Digital
display module will return to
monitoring mode.
Digital Display Module
(Advanced)