9
Part No. 6-847684-00 Rev. D01
Figure 7: Mounting Bracket Adjustment
Rough-In Installation
Pipeline Connection
Figure 8: Remote Sensor Back Box
Sensor Module
1. Braze copper extension tubes from
sensor rough-in to appropriate
pressure/vacuum piping system drops
as shown in Figure 9. Braze
connections per procedures required
by NFPA 99 or CAN/CSA-Z305.1.
Use appropriate measures to prevent
overheating and damage to internal
components of sensor rough-in
assemblies.
2. Perform standing pressure tests and
cross connection tests as required by
NFPA and CSA.
Loosen screws to
adjust back box depth
(same each side)
Four holes in back of box are
provided for surface mounting
Alarm panels ordered with remote
sensors are provided with separate metal
boxes for each sensor. Mount each
sensor box as follows:
1. Remove sensor rough-in from
mounting box (Figure 8).
2. Using four mounting holes provided,
attach box to a wall or other structural
support. (fasteners by others).
3. Reinstall sensor rough-in.
Rear conduit knockouts
(either side) are recommend-
ed for sensor wiring
Alarm Panel Back Box Mounting
Remote Sensor Back Box Mounting