4-5
MEGA3 NFPA Medical Gas Notification System
4107 9016 59.03
Figure 49: Password entry
Enter
Backspace
Close
Clear all
entered
text
Figure 51: Normal Status Message screen
4.2 Alarm Configuration
To set up or change the configuration of an
alarm, start on the CONFIGURATION screen.
Users are always prompted to enter the pass-
word when accessing this screen.
Type the default password (121212), and
touch the Enter button. You will be prompted
to change the password if not done so previ-
ously (Figure 49).
The CONFIGURATION screen displays.
Normal Status Message/Panel ID Descrip-
tion
The MEGA3 alarm allows for customization of
certain labels and messages.
Normal Status Message
To change the message that is communicated
when the alarm is operating in normal status:
1. Touch the [A] button that is located to the
left of the configuration screen (Figure 50).
2. Touch the Erase button on the keyboard
to delete the existing message (Figure 51).
3. Enter the new word or message.
4. Touch the Save button.
Figure 50: Configuration Screen