MEGA3 NFPA Medical Gas Notification System
4107 9016 59.03
Figure 54: Configuring Gas Badges
A = [GAS INPUT]
To change which wired input should be shown on the badge
B = GAS DESC
To change color codes based on standard and country
C = [GAS NAME]
To label the gas
D = AREA MONITORED
To edit the location label for a specific gas
E = [UNITS OF MEASURE]
To change the displayed unit of measure for the pressure/vacuum
F = LOW ALARM SETTINGS
To adjust the value that should indicate that the pressure is low
G = HIGH ALARM SETTINGS
To adjust the value that should indicate that the pressure is high
H = ZERO OFFSET
To make small adjustments to the displayed pressure based on user
preference
Figure 55: Gas Input Tab
Figure 56: Gas Desc Tab
Eight tabs are used for configuration (Figure 54). Tabs are listed in alphabetical order.
A = [Gas Input] Tab
To change which wired sensor should be dis-
played on that badge:
1. Touch the tab for the appropriate wired
gas sensor.
2. Touch the Save button to close the screen.
B = Gas Desc Tab
To change color codes based on your stan-
dard and country:
1. Scroll through the codes using the right
and left arrows at the bottom of the
screen.
2. Touch the tab for the appropriate code.
3. Touch the Save button when you see a
check beside your selection.
Note:
If the alarm default is correct, touch
the X button in the upper right to close the
screen.