MEGA3 NFPA Medical Gas Notification System
4107 9016 59.03
Alarm Type Tab
The ALARM TYPE screen defines which type
of alarm will be displayed and drives later
setup functions.
• Touch the ALARM TYPE tab.
Figure 93: Alarm Type Tab
• Choose a code style based on code re-
quirements.
• Touch the COMBO tab.
• Once the desired tabs are selected, touch
the Save button to save the selection an
go back to the PRIMARY SETTINGS screen.
Note
: To clear your selections, touch the
Erase button.
Initialize Modules Tab
The PRIMARY SETTINGS screen will now dis-
play with ALARM TYPE checked and INITIAL-
IZE MODULES active.
• Touch the INITIALIZE MODULES tab.
The INITIALIZE MODULES screen is used to set
up the boards that are installed. The boards
function as the input/output hardware of the
alarm.
At this point, you have two options:
1. Either touch the Scan button to prompt
the alarm to autofill the screen for you.
OR
2. Complete the screen by selecting the
boards installed in the alarm.
Touch the Save button when you see checks
beside your selections.
Figure 95: Initialize Modules Screen
Figure 94: Choosing Code and Alarm Styles
Save
Erase