102
6-847718-00 Rev. B00
Monitoring Mode (Cont.)
Operation
Multi-Signal modules monitor up to five
dry-contact switches.
• If the signal is normal (switch contacts
closed), green indicator for signal is
illuminated.
• If an alarm occurs (switch contacts
open), audible alarm activates and red
indicator for signal flashes.
• When front panel alarm MUTE
button is pressed, audible alarm is
silenced. Alarm indicator remains
illuminated, but does not flash.
• Visual indicator will remain illuminated
as long as alarm condition remains.
• If audible alarm repeat time has been
set, audible alarm will reactivate, and
visual indicator will again flash after
specified time.
Signal 1
Signal 2
Signal 3
Signal 4
Signal 5
Multi-Signal Module
Figure 96: Multiplexer Module