114
6-847718-00 Rev. B00
Periodic testing of alarm system is
recommended. Alarm modules have built-
in self-test modes.
Front panel TEST button will initiate self-
test mode on all alarm panel modules at
same time.
Front panel
TEST button
Press and hold the front panel TEST but-
ton for 5 s to initiate the following
sequence:
•
Audible Alarm Test
Audible alarm sounds and VFD
displays for 1 s.
•
LED Test
VFD displays for 9 s while other
modules perform LED tests.
•
VFD Test
VFD scrolls the following message on
both rows
Test ABCDEFGHIJKLMNOPQRSTUVWXYZ
Test ABCDEFGHIJKLMNOPQRSTUVWXYZ
•
Checksum and IP Address Display
VFD displays checksum and device IP
address for 5 s.
C
= 4 hex digit checksum
S
=
Ok
if checksum is correct or
Bad
if checksum is incorrect
X
= IP Address
•
Signal Display
VFD displays each signal parameter for
2 s.
If the gas type for the signal is
Unused
the input gas type is displayed for 0.5 s.
Input XXX
LED D
Input XXX Location
LLLLLLLLLL
Input XXX Message
MMMMMMMMMMMMMMMMMMMM
Input XXX Gas Type
GGGGGGGGGGGGGGG
Checksum: CCCC SSS
IP: XXX.XXX.XXX.XXX
LED Test
Audible Test
Annunciator Module
Alarm System
Testing
Figure 108: Alarm Panel Test button
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...