27
6-847718-00 Rev. B00
• Power all alarms from life safety
branch of emergency power system as
required by applicable standards.
• Protect all wiring from physical
damage by raceways or conduit as
required by applicable standards.
• Wire master panels directly to
switches or sensors as required by
applicable standards.
• Wiring runs should be made with color
coded wire. Record color, signal, and
source of signal for each wiring lead to
aid in connection of alarm finish
components.
• Avoid installing alarm panels near
radio transmitters, electrical motors, or
switchgear.
All low voltage wiring must meet the fol-
lowing criteria:
• Copper wire no smaller than 22 AWG.
• Circuit length not to exceed 5000 feet.
• Cable must be twisted-pair shielded
type. Multi-pair cables within one
common shield are acceptable.
Wire Type And Size
General Requirements
Pull Wire
Inst
allation
Procedures
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...