28
6-847718-00 Rev. B00
Holes for entrance of
low voltage field wiring
Two choices of
conduit holes for
entrance of AC
mains power
top entry
side
entry
For top entry, remove
shield for access.
Reinstall shield after
connection is made
Figure 22: Wire Routing for Area Alarms
Figure 21: Wire Routing for Master / Combo Alarms
Two holes or knockouts are provided for
entrance of AC mains power. One hole
on left side and one hole on top left of
alarm panel back box.
NO OTHER HOLES SHOULD BE
PUNCHED OR USED.
Several additional holes or knockouts are
provided on right side of top panel and
bottom for entrance of low voltage field
wiring(Figures 21 and 22).
Two choices of
conduit knockouts
for entrance of AC
mains power
top
entry
side
entry
Knockouts for entrance of
low voltage field wiring
Knockouts for entrance of
low voltage field wiring
Wire Routing
Pull Wire (Cont.)
Inst
allation
Procedures
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...