59
6-847718-00 Rev. B00
Operation
Annunciator Module
Set-Up Procedures
Setup Using Front Panel Controls
(Master and combo alarm panels)
Many parameters may be set using front
panel controls, but some parameters must
be set up using web pages.
1. Press SETUP button on back of
annunciator module to enter setup
mode.
2. N2 Address
This parameter is the Johnson Controls
Metasys® N2 Address.
Possible values are:
1
-
255
(Factory default:
1
)
3. Clear Network
Answering ‘Yes’ will clear and rebuild
the medical gas network.
4. Reset Users
Answering
Yes
will clear all user ids
and reset User 1 name and password
to factory defaults:
Username: new
Password: new
Reset users
[YES/NO]
Clear network
[YES/NO]
N2 Address
[VALUE]
Press or to
change the value
Press ‘TEST’ for
the next item
**** Device ****
**** Setup ****
NOTE:
Before setting up signals, fill out Master Alarm
Signal Input Data table on page 138.
5. Input gas type
Sets the gas type for signal
XXX
.
Possible GAS TYPES are:
Oxygen
Nitrous Oxide
Medical Air
Vacuum
Nitrogen
Lab Air
Lab Vacuum
Argon
Instrument Air
WAGD
Carbon Dioxide
CO2-O2
O2-CO2
He-O2
O2-He
Helium
Custom
Unused
Dental Air
If gas type is set to ‘Unused’, step 5 is
repeated for the next input number.
6. Input Message
Sets the alarm message for signal
XXX
.
There are several pre-programmed
messages or a custom message can
be selected. A custom message must
be entered using web page setup.
Possible values depend on gas type
selected in previous step (See Table 1).
7. Input lights LED
Assigns LED to signal
XXX
.
Possible values are
1
-
10
for master
and large combo alarms, and
1
-
5
for
small combo alarms.
Input XXX lights
LED
[LED Number]
Input XXX message
[Message]
Input XXX gas type
[GAS TYPE]
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...