65
6-847718-00 Rev. B00
Operation
Set-Up Procedures (Cont.)
0
0
to
11 33 8
8 0
0
(
11 77 2
2 55
*
) (kPa)
0
0
to
2
2 0
0 0
0
(
2
2 55 0
0
*
)
(psig)
0
0
to
33 2
2
(in Hg)
0
0
to
8
8 0
0 0
0
(mm Hg)
*N2 and instrument air only
6. Press and release TEST button.
HIGH indicator will illuminate and
K
K ,,
will be displayed. After two seconds
current high alarm setting will be
displayed. Select desired high alarm
setting by pressing either
!
or
"
buttons. High alarm setting ranges
are:
0
0
to
11 33 8
8 0
0
(
11 77 2
2 55
*
) (kPa)
0
0
to
2
2 0
0 0
0
(
2
2 55 0
0
*
)
(psig)
0
0
to
33 2
2
or
NN OO NN EE
(in Hg)
0
0
to
8
8 0
0 0
0
or
NN OO NN EE
(mm Hg)
*N2 and instrument air only
7. Press and release TEST button to
save changes.
D
D OO NN EE
will be
displayed for two seconds.
8. Digital display module will return to
monitoring mode.
NOTE:
If module is idle for more than one minute dur-
ing set-up procedure, module will chirp three
times, return to monitoring mode and will
default to previous settings. No changes to
settings will be saved.
The set-up must be completed and the module
must display
DD OO NN EE
in order for changes to be
saved.
Digital Display Module (Cont.)
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...