66
6-847718-00 Rev. B00
Operation
Digital Display Module-
Advanced
Two additional levels of “advanced” set-up
features are available.
The first advanced level,
SS EE TT 2
2
, is used to
adjust communication related parameters.
Changing Communication Parameters
A
A D
D D
D RR
(network address)
A network address uniquely identifies each
digital display module in the alarm panel.
Address range is
0
0
to
2
2 55 55
.
For TotalAlert 2 alarm panels, this setting is
programmed at the factory when the alarm
panel is assembled and should not be
changed.
BB A
A U
U D
D
(baud rate)
Used to set communication speed. User
choices are
44 8
8 0
0 0
0
or
99 66 0
0 0
0
.
Baud rate must be set to
99 66 0
0 0
0
. for
TotalAlert 2 alarm panels.
NN D
D FF
(network data format)
Defines data word length. Choices are
8
8
(bits) or
11 66
(bits). NDF must be set to
11 66
for TotalAlert 2 alarm panels.
P
P D
D FF
(pressure data format)
Defines data word length transmitted to a
slave alarm panel. User choices are
8
8
(bits) or
11 66
(bits). Factory setting is
11 66
.
Set-Up Procedures (Cont.)
1. Press and HOLD SETUP button for
TWO seconds until
SS EE TT 2
2
is displayed.
2. Press and release TEST button on
front of alarm panel.
A
A D
D D
D RR
will be
displayed for one second and then
current network address. Select
desired address by pressing
!
or
"
buttons.
Each module must have a unique
address.
3. Press and release TEST button.
BB A
A U
U D
D
will be displayed for one second
and then current baud rate. Select
desired baud rate by pressing
!
or
"
buttons.
4. Press and release TEST button.
NN D
D FF
will be displayed for one second and
then current network data format.
Select desired data format by
pressing
!
or
"
buttons.
5. Press and release TEST button.
P
P D
D FF
will be displayed for one second and
then current pressure data format.
Select desired data format by
pressing
!
or
"
buttons.
6. Press and release TEST button to
save changes.
D
D OO NN EE
will be
displayed for two seconds.
7. Digital display module will return to
monitoring mode.
NOTE:
If module is idle for more than one minute dur-
ing set-up procedure, module will chirp three
times, return to monitoring mode and will
default to previous settings. No changes to
settings will be saved.
Set-up must be completed and module must
display
DD OO NN EE
in order for changes to be
saved.
CAUTION:
Communication parameters are factory set and
should not be changed. These instructions are
provided for programming replacement modules.
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...