Operation
Set-Up Procedures (Cont.)
Digital Display Module-
Advanced
(Cont.)
The second advanced level,
SS EE TT 33
, is
used to adjust other miscellaneous
parameters as listed below:
Changing Miscellaneous Parameters
SS EE RR U
U
(gas service identification)
Used to set digital display module’s gas
service identification for communication
with a digital sensor.
User choices are:
0
0 2
2
(oxygen)
NN 2
2 0
0
(nitrous oxide)
A
A ,, RR
(medical air)
U
U A
A ((
(vacuum)
NN 2
2
(nitrogen)
LL A
A ,, RR
(lab air)
LL U
U A
A ((
(lab vacuum)
A
A RR 99 NN
(argon)
II A
A ,, RR
(instrument air)
U
U A
A 99 D
D
(WAGD)
(( 0
0 2
2
(carbon dioxide)
(( -- 0
0 2
2
(CO2/O2)
0
0 2
2 -- ((
(O2/CO2)
K
K EE 0
0 2
2
(He/O2)
0
0 2
2 K
K EE
(O2/He)
K
K EE
(Helium)
NN OO NN EE
(none)
OO FF FF SS
(display offset)
Used to adjust digital module’s display.
User choices are
0
0
,
11
,
2
2
,
-- 11
, or
-- 2
2
.
Factory setting is
0
0
.
RR EE P
P TT
(audible alarm repeat)
Used to set time interval for reactivation of
audible alarm after alarm MUTE
button
has been pressed.
User choices are:
OO FF FF
11
,
55
,
11 55
,
33 0
0
(minutes),
11 HH RR
,
8
8 HH RR
,
11 2
2 HH RR
or
2
2 44 HH RR
(hour).
Factory setting is
OO FF FF
.
When set to
OO FF FF
, audible alarm will not
reactivate after an active alarm has been
silenced. Any new alarm will reactivate
audible alarm.
II NN TT SS
(display intensity)
Used to adjust digital module’s display
brightness.
Factory setting is maximum intensity.
1. Press and HOLD SETUP button for
FOUR seconds until
SS EE TT 33
is displayed.
2. Press and release TEST button on front
of alarm panel.
SS EE RR U
U
will be displayed
for one second and then current gas
service identification. Select desired
gas service by pressing
!
or
"
buttons.
3. Press and release TEST button.
OO FF FF SS
will be displayed for one second and
then current display offset. Select
offset by pressing
!
or
"
buttons.
4. Press and release TEST button.
RR EE P
P TT
will be displayed for one second and
then current audible alarm repeat time.
Select desired repeat time by pressing
!
or
"
buttons.
5. Press and release TEST button.
II NN TT SS
will be displayed. Select desired
display brightness by pressing
!
or
"
buttons. Each push will change
brightness of display. Audible alarm will
chirp one time with each push and two
times when limit reached.
6. Press and release TEST button to save
changes.
D
D OO NN EE
will be displayed for
two seconds.
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...