78
6-847718-00 Rev. B00
Operation
Each TotalAlert 2 alarm panel keeps a list
of up to 75 other devices on the medical
gas network.
1. Click on Enroll Devices to access the
Enroll Devices Setup page (Figure 59).
2. Click on an unused Device # link to
enroll a new device.
3. Enter the device IP address and Device
Serial Number then click Submit
(Figure 60).
4. The Enroll/Delete Device Complete
page will be displayed (Figure 61).
5. Click Return to Enroll Devices to return
the the Enroll Device Setup page.
Enroll Devices
Set-Up Using Web Pages (Cont.)
Figure 59: Enroll Device Setup
NOTE:
The factory default settings will cause all devices
on the same subnet to be automatically
discovered. Enroll Devices is only necessary if
the alarm network is spread across multiple
subnets.
NOTE:
Device # links are disabled for automatically
enrolled devices.
NOTE:
Fixed IP addresses should be used for manually
enrolled devices. If a manually enrolled device’s
IP address changed, the entry in the device list is
NOT automatically updated.
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...