86
6-847718-00 Rev. B00
Operation
Setup Device
Set-Up Using Web Pages (Cont.)
This setup page is used to configure alarm
panel name, location, N2 address, facility
description and contact information.
1. Click Setup Device to access the Setup
Device page (Figure 72).
2. Enter the new device name.
To access the web pages by using the
name you have given it, NetBIOS
Name Service over TCP/IP must be
enabled on your computer (See Set-Up
Network Interface on page 72).
• Without the NetBIOS Name Service
enabled, the device must be
accessed by using its IP address.
• To learn a master or combo alarm’s
IP address, press and hold the
TEST button for 5s.
3. Enter the Location.
4. If the device will be connected to a
Johnson Controls Metasys® building
automation system, select the N2
address.
5. Enter the Facility Description.
6. Enter the Contact Information.
7. Click the Submit button.
8. The Device Setup Accepted page will
be displayed (Figure 73).
NOTE:
The information in these fields identify the alarm
panel that generated an alarm or an event. This
information is displayed on the event log and on
electronic notification messages.
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...