6.5 General
Inspections
6.5.1 Monthly
Inspections
A general inspection should be performed on a
regular basis (monthly) for safety items. Items to
inspect include all wiring,
fl
ex hoses, and other
items. If a damaged item is viewed, call your local
BeaconMedæs
service technician for a thorough
inspection and report of
fi
ndings
6.5.2 Every Six Months
A thorough inspection of the vacuum pump
cooling air intake and discharge grating should
be performed at least every six months or more
frequently if conditions require. If a dust/dirt
buildup is visible, clean the grating to remove the
buildup. Excess dust/dirt buildup in these areas
will prevent air from cooling the vacuum pump,
a
ff
ecting performance of the vacuum system.
6.5.3 Oil
Leak
If upon inspection of the vacuum system, a
noticeable amount of oil is seen on the base of
the system or on the
fl
oor around the system,
immediate action is required to prevent the
possibility of injury. Using an oil absorbent or
like product, absorb the oil on the base frame
or
fl
oor to remove the potential of slipping. Call
your local
BeaconMedæs
service technician for
a thorough inspection of the vacuum pumps.
gasket in place, secure the exhaust cover end
plate (140) to the exhaust box using eight hex
head cap screws (142) and eight lock washers
(143).
For 15 Hp, 20 Hp, and 25 Hp
To replace the exhaust
fi
lters, unscrew the
screws (142) and lock washers (143) from the
exhaust cover plate (140). Remove cover plate
and gasket (141).
Use a slotted head screw driver to loosen the
exhaust
fi
lter retaining spring (125), then rotate
and remove the spring. Pull the
fi
lter cartridge
(120) out of the exhaust box.
Reinstall the
fi
lter elements. Make sure the open
end of the element is properly seated down in
its recess in the exhaust box with the O-ring
(121) correctly positioned. The indicating arrow
on each
fi
lter element must be pointing upward
toward the top of the exhaust box after the
assembly has been installed.
Retain the
fi
lter with the spring clip (125), tighten
the tension screw until the
fi
lter is secure.
Place the exhaust port gasket (141) and cover
(140) in position on the exhaust box and retain
with the cap screws (142) and lock washers (143).
6.4 Receiver Tank Drain
The receiver tank drain is located on the extreme
bottom of the tank and is used to drain o
ff
fl
uid
accumulation. The valve can only be used when
the tank is isolated and the tank is at atmospheric
pressure.
NOTE:
Refer to Figure 6.5 for numbers referenced
below in parenthesis.
Lubricated Rotary Vane Laboratory Vacuum Systems
4107 9021 88.01
6-6