pages. The maximum number of pages is 200
(1000 events divided by 5 events per page).
• Page up and Page down arrows change the
view 5 events up or down at a time and are
only visible if subsequent event information
is available. The Home button changes the
view to the top or most recent events and
the End button changes to the bottom or the
oldest events.
• A maximum of 1000 most recent events are
stored for the system.
A.2.11 5.7” Settings Screen
The settings screen (Figure A.13) allows the
selection of various sub screens that pertain to
system con
fi
guration data. All value/adjustment
changes are password protected.
• Vacuum – Allows the adjustment of system
vacuum operating range.
• Audible – Allows the horn to re-initiate if an
alarm or shutdown event has not cleared,
but the horn was silenced. The allowable
values for re-initiate time are never, 15 min,
30 min, 1 hour, 8 hours, and 24 hours.
• Language – Allows the choice of pre-de
fi
ned
display language.
• Units – Allows the display units to be displayed
and changed.
• Model – Displays the pertinent system model
information.
• Date/Time – Allows the date/time to be
displayed and changed.
• Network – Allows the network (IP address,
etc) to be displayed and changed.
• General – Allows the miscellaneous general
information to be displayed and changed.
Figure A.14 Lab Vacuum Settings Screen
Figure A.12 History Screen
Figure A.13 Settings Screen
General Settings Screen
Lab Vacuum Settings Screen
Lubricated Rotary Vane Laboratory Vacuum Systems
4107 9021 88.01
A-7