1. Click System Trend Setup to access the Sys-
tem Trend Setup page.
2. Select a time interval from the pull-down list
of options.
3. Click the Submit button.
A.7.10 Electronic Noti
fi
cation Setup
This Electronic Noti
fi
cation Setup page (Figure
A.36) is used to con
fi
gure the Electronic Noti
fi
ca-
tion feature of the TotalAlert Embedded control
system. By setting up the Electronic Noti
fi
cation,
key personnel can receive noti
fi
cations of all
alarm/shutdown alerts and/or all maintenance
alerts (e-mail, pager, phone text message). The
device acts as an SMTP client. An SMTP server is
required for electronic noti
fi
cation to function.
1. Click Electronic Noti
fi
cation Setup to access
the Electronic Noti
fi
cation Setup page.
2. Select Enable to enable the e-mail noti
fi
ca-
tion tool.
3. Enter up to
fi
ve email addresses.
4. For each address, select “System” for that
person to receive all alarm/shutdown alerts.
Select “Maintenance” for that person to re-
ceive all maintenance alerts. An individual
may receive both types of alerts.
5. Enter an email address in the “From” box as
this will be the sender of the noti
fi
cations.
6. If desired, early warning noti
fi
cations can
be set up for maintenance items. By select-
ing these time intervals, a noti
fi
cation will
be sent before the maintenance item is due,
providing time for planning the service activ-
ity.
7. Click the Submit button.
A.7.11 Navigating the Website
The TotalAlert Embedded website allows the user
to easily view the status of all activity pertaining
to the laboratory vacuum system. By clicking the
menu items to the left of the screen, a user can
view pages displaying accurate and timely infor-
mation about the system. These pages include:
Status:
The Status page (Figure A.37) displays
current information for both the overall system
(run hours, vacuum level, ambient tempera-
ture, alarms, shutdown, and service) and the
units (run hours, operation mode, run mode, se-
quence, alarms, shutdown, and service).
Figure A.36 Electronic Noti
fi
cation Setup
Figure A.37 Status Page
Figure A.38 Alarms Page
Lubricated Rotary Vane Laboratory Vacuum Systems
4107 9021 88.01
A-17