The connecaon of the sucaon tube ⑩ is firmly connected with the cover ⑪. To remove
the whole cover of the filter chamber ⑪ draw gently the connecaon part of the sucaon
tube ⑩. Then hold the filter bag with your hand at the front, to avoid the coming out of
the dust, and pull the hose flange from the opening. Remove the full filter and replace it
by a new one. The replacement and re-assemble following the above steps in reverse
order
3. T
EMPERATURE
PROTECTION
OF
THE
SUCTION
TURBINE
If the air temperature in the sucaon turbine will exceed 80° C. the turbine and the
handpiece will be switched off. On the display the error message
will appear.
A|er falling of the temperature inside the case of the sucaon turbine of approx. 10 - to
15 °C this announcement will disappear and then the device will be in stand-by condiaon.
TIP:
The standard commercial smell cartridges for domesac vacuum cleaners
(accessories) fit in the filter bag opening and prevent extensively the smell
nuisance.
N
OTE
:
P
OSSIBLE
REASONS
FOR
AN
OVERHEATING
OF
THE
SUCTION
TURBINE
ARE
:
-
Blockage in sucaon hose or sucaon handpiece
(refer to chapter VI, point 1 „Cleaning of the sucVon handpiece“, page 15)
-
Overfilled filter bag
(refer to chapter IV, point 3 „Filter change“, page 12)
-
Insufficient air circulaaon at the rear of the unit
(refer to chapter I, points 5-6, page 5)
If the above menaoned reasons can be excluded, we recommend to contact an
authorized dealer for repair or maintanance services.
9