1
Specifications and
Preinstallation Requirements
1-1
SPECIFICATIONS
Only values with tolerances or limits are guaranteed data. Values
without tolerances are informative data, without guarantee.
CONTROL FEATURES
Speed
Setting range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 to 26 000 rpm
(in 100-rpm increments below 10 000 rpm,
500-rpm increments above 10 000 rpm),
or
equivalent RCF (in 100
×
g
increments)
Elutriation speed setting range . . . . . . . . . . . . . . . . . . . . 0 to 5 000 rpm
(in 10-rpm or 10
×
g
increments)
Speed display . . . . . . . . . . . . . from 0 to 10 000 rpm, display indicates
actual rotor speed ±10 rpm
from 10 000 to 26 000 rpm, indicates actual
rotor speed ± 0.1%, or equivalent RCF
Time
Setting range . . . . . . . . . . . . . . . . . . . to 180 min or continuous (hold)
Time display. . . . . . . . . . . . . indicates run time remaining (timed run)
or elapsed time (hold run)
Temperature
Setting range . . . . . . . . . . . . . . . . . –10 to +40°C (in 1°C increments)
Accuracy . . . . . . . . . . . . rotor temperature controlled to within ±2°C
of set temperature (after equilibration)*
Ambient temperature range . . . . . . . . . . . . . 16 to 38°C (60 to 100°F)
†
Cooling fluid . . . . . . . . . . . . . . . . . . . . . . . . . Refrigerant 404A (HFC)
Acceleration maximum or slow
Deceleration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . maximum, slow, or off
______________
* During transient conditions, such as acceleration and deceleration, rotor temperature may be outside this range. Refer to applicable rotor manuals
for specific rotor operating range information.
†
To reach temperatures above ambient, the centrifuge is dependent on the frictional heat generated inside the chamber during operation. At low
run speeds or low ambient temperatures, the centrifuge may not be able to achieve some higher temperatures.