A-2
Temperature Calibration Procedure
5. After at least 1 hour, measure the temperature of the buffer or
water using a thermometer or thermocouple.
6. If the measured and set temperatures are different, note how many
degrees different they are and adjust the set temperature up or
down that number of degrees.
For example:
If the required sample
temperature is
And the measured
buffer/water
temperature is
Set the temperature
to
4°C
6°C
2°C
5°C
4°C
6°C