background image

MD3150/3060 

 

 

Copyright BEGLE GmbH, All rights reserved.  

Page 7 

Internet: www.bue.de 

 

 

Mode 2: 

 

Set time interval  

 

 

 

Mode display 

Small display 

Large display 

 

 

 

Time interval in seconds 

 

 

 

  The measuring interval can be set between 1 and 9999 seconds. At the end of the 

set interval the next measured value is stored in the memory automatically (max. 800 
measured values). 
Pressing the plus key increments the time interval, the minus key decrements it. 
Keeping a key pressed automatically increments / decrements the time interval (auto-
increment). Time intervals > = 64 sec. enable the power save mode when a 
measuring series is started in mode 3. 

 

Mode 3: 

 

Interval measurement 

 

 

 

Mode display 

Small display 

Large display 

 

 

End,StP,memory loc. no. 

Current temperature, HLt 

 

 

 

 

 
 
 

 
 
 
 
 
 
 
 
 

 

  If no interval measurement is active, the 

small displays

 shows ‘End’ and the 

large 

display

 shows the current temperature value. 

 
Pressing the plus key starts an automatic interval measuring mode provided an 
interval timer > 0 has been entered in mode 2. 
The 

small display

 always shows the current memory location number whereas the 

large display 

shows the temperature value for that location. 

Pressing the minus key stops the interval measurement. In this case the 

small display

 

shows ‘StP’. Pressing the plus key again continues the measurement in intervals. If 
the measured value memory is full (800 values), the interval measurement stops 
automatically. The active interval measurement also stops when the minus key is 
pressed twice. The 

small display

 shows ´End´. 

If a trigger function is active in mode 7 and an interval measurement is started, bars 
are displayed in the top or bottom segment row (

small display

) as long as the trigger 

temperature (mode 6) has not been reached. The actual interval measurement starts 
only when the trigger temperature set in mode 6 is reached (actual temperature value 
above or below the trigger). 
If the mode is changed while an interval measurement is active, mode 4, mode 8, 
mode 0 and mode 1 are not available. If the interval measurement is interrupted, 
mode 8 and mode 0 are skipped. The min, max and mean value memories are 
deleted after the interval measurement is started.  
 
Power save mode: 
If a time interval of > 64 seconds is set in mode 2, the measuring instrument changes 
to power save mode after the measured value is stored. The 

large display

 shows 

´HLt´. The measuring instrument is activated again just before the next measurement 
and returns to power save mode again after the measurement 

(power consumption in 

power save mode is approximately 1mA)

 
The power save mode (´HLt´) can be exited by holding the minus key pressed. This 
also automatically stops the interval measuring mode. 
During long-term measurement the series can only be stopped (

small display

 = ‘StP’) 

at the time the measured value is saved in the memory. 

 

Summary of Contents for MD3060

Page 1: ...ant note Please read Please read this Instruction Manual attentively Damage due to non compliance with the instructions in this Manual is not covered by guarantee We are not responsible for any consequential damage which is the result of this ...

Page 2: ... been filed with the manufacturer To maintain the thermometer in this state and ensure its safe use you must follow the instructions in this manual If you have any questions you can find more information tips and updated manuals on our website www bue de This symbol precedes important information which should be noted Imprint This Instruction Manual is published by Beckmann Egle GmbH Kernen All ri...

Page 3: ...that you use the thermometer only if the measuring media are electrically isolated The user is obliged to ascertain whether the thermometer is suitable for the purpose for which he intends to use it Do not use the thermometer in adverse ambient conditions Examples of adverse ambient conditions are wet or high relative humidity dust or combustible gases vapors or solvents thunderstorm or similar co...

Page 4: ...mometer can only be switched off in this state Please replace the battery without delay Spent batteries must be disposed as hazardous waste Connecting the MD3150 The temperature measuring instrument has two different connections for 8 pin MINI DIN plug This connection can be found on the front of the temperature measuring instrument see figure 9V block battery battery compartment on rear side Conn...

Page 5: ...y to switch on the device Mode functions Press the Mode key a the keys as needed Switching off Press the ON OFF key once more The following data is restored when the device is switched on Measured values of the series Min Max mean value storage State settings Intermittent measurement C F Time controlled switch off on off Interface output on off Brief use guide Mode 0 Changes the channel 1 2 1 2 on...

Page 6: ... channel 2 C12 temperature display alternates between both channels every 2 seconds Press the minus key to change the temperature display between units of C and F Mode 1 Switching the serial interface on off output of measured values in the memory Mode display Small display Large display 1 ON or OFF If the small display is ON briefly pressing the minus key causes output of the measured values in t...

Page 7: ...tervals If the measured value memory is full 800 values the interval measurement stops automatically The active interval measurement also stops when the minus key is pressed twice The small display shows End If a trigger function is active in mode 7 and an interval measurement is started bars are displayed in the top or bottom segment row small display as long as the trigger temperature mode 6 has...

Page 8: ...display Large display 5 0 Min 1 Mean 2 Max Tmin Tmean Tmax Pressing the plus key shows 0 1 2 in the small display The large display shows the appropriate minimum mean and maximum temperature values These memory entries are automatically deleted when the next interval measurement is started mode 3 The above 3 values can be deleted manually with the minus key In this case the values 20 C 0 C 120 C M...

Page 9: ... off automatically Mode 8 Manual save Mode display Small display Large display 8 Memory location no Current temperature The current temperature and the memory location number are displayed To carry out a series of measurements in manual save mode a memory reset to 000 must at first be made with the minus key Each press of the plus key saves the current temperature value at the indicated memory loc...

Page 10: ...Please observe the following instructions Spent batteries are hazardous waste and must be disposed as prescribed by law Do not use carbon containing chemicals or fuel alcohol or similar fluids for cleaning Damage due to non compliance with the above instructions is not covered by guarantee Troubleshooting and remedy Problem Possible remedy The device does not display temperature but only meaningle...

Page 11: ...minal under Windows settings 1200 n 8 1 handshake hardware or RTS ON Switch on the measuring instrument and ON in the small display in mode 1 Press the mode key 7 times to enter mode 8 The current temperature value is displayed The interface remains active even if data is not sent When the mode key is pressed again the temperature values are output on line to the interface in the above data format...

Page 12: ...acters MD3150 temperature sensor pin configuration Plug soldered end 1 2 3 4 5 6 7 8 NTC 1 NTC 2 Sensor cable 1 2 3 4 5 6 7 8 Channel1 Channel2 Common channel Surface sensor Connects to channel 1 and channel 2 Pierce sensor Connects to channel 1 Y adapter If connection is not marked channel 1 and channel 2 remain connected 1 1 if marked red channel 1 and channel 2 have been mixed up internally con...

Page 13: ...ng parameters copying the measuring curve and measuring parameters in the Windows clipboard for further processing by other Windows programs Functions On line monitoring of the temperature on the PC monitor screen Freely selectable temperature range and number of measuring values Saving the measuring values and measuring parameters in ASCII text file format Transfer of the measured values stored i...

Page 14: ... Page 14 Internet www bue de Connecting cable PC interface cable MD3142 Y adapter MD3143 mains adapter MD3144 MD3150 special accessories 1 channel pierce sensor MD3131 2 channel surface sensor MD3123 Plug with sensor cable for MD3121 Separate sensor beads MD3120 ...

Page 15: ...or Surface sensor air or gas cooler Sensor shape 2 Feeler directly with thermo plug Dip sensor pierce sensor Surface sensor air cooler Sensor shape 3 Cable sensor for quick measurement response time 0 3 s MD3024 65 to 300 C MD3034 25 to 400 C Thermo extension cable MD3045 Föat tip Cone tip Thermocouple tip Spring mounted thermocouple tip Copper plate at tip Fat tip Cone tip Thermocouple tip ...

Page 16: ... mm Time trigger Interval 1 to 9999 seconds Power save mode for long time measurement t 64 s Manual trigger One value measured at any time interval Automatic threshold trigger Tist trigger temperature or Tist trigger temperature Display Temperature 4 digits either C or F Function 3 digit Mode 1 digit MD3150 Measuring range 20 0 to 120 0 C at 0 1 C resolution Accuracy Maximum deviation 0 1 C deviat...

Reviews: