Basic principles: Crucible
Adjusting the crucible locking device
The crucible in use can be locked in its position by
means of the crucible locking device, thus ensuring
that the melt is safely drained by the centrifugal
force.
1.
Loosen the wing screw on the crucible lock.
The crucible locking device can now be
adjusted.
2.
Pull die crucible locking device backwards
using the wing screw.
3.
Insert and align the crucible so that the outlet
of the crucible fits into the opening provided for
that purpose.
4.
Push the wing screw back towards the crucible
so that the crucible locking device is locked in
its position.
5.
Tighten the wing screw.
Recommendation: Only use BEGO crucibles (REF
52483 - identifiable by the BEGO logo).
WARNING
Burn hazard from hot parts!
• Always use mould tongs to move
hot crucibles.
• Always use mould tongs to move
hot crucible inserts!
NOTICE
Fornax ceramic crucibles (RE 52483) do not
need to be preheated before casting.