Manual casting without pyrometer
The symbol "M" is shown in the display header.
With the NAUTILUS
®
CC plus, manual casting
without pyrometer is used as an optional
"emergency operation" in case the pyrometer is
malfunctioning/not available.
1.
Go to "Settings" > "Program".
2.
Deactivate the settings (pyrometer and auto-
matic casting), the "Manual casting" button is
shown in yellow.
3.
Press "Back" to return to the casting process.
4.
Select the alloy (see page 42).
5.
Optional: Enter log data (see "Setting up the
protocol" on page 41). This data appears in the
casting log. Casting logs can be exported to a
USB flash drive with the "Export casting logs"
button.
• Enter work number, batch and casting quan-
tity and confirm the entry with "OK" or
• Skip the data entry with "Abort".
6.
Insert the crucible and alloy.
7.
Close the crucible door.
8.
Confirm the preheating process with "OK".
In fast mode, the preheating process starts
immediately when the crucible cover is closed.
ATTENTION
The temperature cannot be
measured and therefore cannot be
limited without the pyrometer!
Stay with the device and finish the
preheating and melting as described
manually!
Summary of Contents for 39701001
Page 1: ...Gerätedokumentation Miteinander zum Erfolg ...
Page 2: ......
Page 4: ......
Page 11: ...English Nautilus CC plus Nautilus T 7 86105 BA en 01 en Unpacking Packing REF 17909 ...
Page 47: ...Nautilus CC plus en English ...
Page 63: ...Nautilus T en English ...
Page 87: ...English Nautilus CC plus Nautilus T 83 86105 BA en 01 en NAUTILUS T bar ...
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......