34
85
95
5 B
A
-en
/1
1
English
⋅
Eltropol 300
en
7.
As a general rule, polishing must be effected in
two stages. The polishing object must be posi-
tioned anew in the bath each time (rotate
180°).
8.
If the unit is not needed for longer periods,
switch it off.
Notes
•
Polishing can start before the operating temper-
ature is reached. After 15 minutes, the Wirolyt
has already reached approx 30 °C. The bath
continues to heat up during polishing.
•
The necessary polishing time and current
strength are both calculated by the unit. These
are influenced by the size of the object and the
bath temperature. If required, the settings may
be adjusted manually:
a) The polishing time may be set anywhere
between 60 seconds (00:60) and
12 minutes (12:00).
b) During polishing the current strength is indi-
cated by the LEDs (ammeter function). In
the default setting (level 3 = 100%) the cur-
rent may attain a maximum of 10 A, how-
ever this is only the case when polishing
two large objects in a hot bath. The current
may be manually limited to 70 % (level 2) or
50 % (level 1) of the calculated current. The
number of LEDs illuminated changes
according to level. After each polishing the
unit automatically returns to level 3!
•
The polishing process can be interrupted by
pressing the Start button. Press again to call up
the polishing time calculated at the outset (not
the remaining polishing time). This may now be
altered manually. Press the Start button again
to start polishing with this setting.
•
Uninterrupted polishing may cause the permit-
ted bath temperature to be exceeded. The unit
then displays the error message "E 30". Having
switched off the unit, wait approx. 1 hour before
switching it on again to resume polishing.
•
An integrated fan switches on automatically, as
necessary, to cool down the unit electronics.
•
WARNING
If the object is to be polished a second
time, allow to dry thoroughly!
Never add water to acid!