Additional information
E01 – E03 are created, when the sensor signal is< 1° C or > 120° C. The errors switch off
the motor and the heaters (safety). The device cannot be used any longer.
E04 is generated by the internal rotation sensor of the motor. This error is appearing alter-
nately with the gel temperature in the display (Emergency operation).
E05 is generated when the timer for the required clock doesn’t run. This bug switches off
the motor and heaters, the device cannot be used any longer. Power failures during timer
operation are intercepted by a maintenance-free storage capacitor.
E20 is generated when the 24V power supply for the motor has failed. This bug switches off
the motor and heaters, the device cannot be used any longer.
E3x is generated when the cover is opened while the motor is running. The cover switch
controls a relay and a brake of the motor. If the motor doesn’t stop immediately , the brak-
ing function hasfailed. The emergency operation is possible, but strongly discouraged.
Keep the cover closed!
Note: the previously used error E8 (cover open) was replaced by the word ‘OPEN’ in the
display.