en
Service menu USER
Fig. 9
Service menu - User level
Sysinfo
The [ Get ] button opens the following window to
indicate system information.
Fig. 10
Sysinfo
Maintenance recommendation
Basics
The system automatically generates and displays
maintenance recommendations for the events
displayed in the window below.
The maintenance recommendation messages are
displayed not immediately after having reached the
respective threshold but only after the following
conditions:
•
ECO-Modus has been finished, or
•
system has been switched on, or
•
six hours have passed since a condition for a
maintenance recommendation is complied with
without any of the above mentioned actions
have occured.
The indicated values represent the percentage for
the respective events in relation to the defined thres-
holds (for details see the following description).
The [ Get ] button of the Sysinfo window opens the
following window.
Fig. 11
Maintenance recommendation
Extraction filter degeneration
This feature is not used.
Water filter degeneration
This relative value refers to the conductivity measu-
rement of the deionized water of the lamp cooling
circuit.
Lamp degeneration
This value is the relative number of laser pulses
emitted since the last lamp counter reset relative to
the setting of the Lamp warn level in the Lamp
settings dialog of the
→
SERVICE MENU
ENGINEER.
Safety check
Elapsed time relative to the fix predetermined check
interval.