24
860
76
D
B
-en/
03
35
5.0
00
32
English
⋅
LaserStar T Plus
en
Installation
This chapter contains a description of the prerequi-
sites for trouble-free operation of the system, as well
as instructions for setting up, starting up and trans-
porting the device.
Requirements
To guarantee a faultless operation of the device the
following requirements have to be met.
Installation site
The installation site has to meet the following
requirements:
•
The device has to be installed in a room as dry
and dust-free as possible.
•
Do not expose the device to direct sunlight.
•
It is recommended to set up the device so that,
if needed (cleaning the grid of the heat
exchanger, checking the coolant water level),
the two sides and the back of the device are
easily accessible.
Ambient conditions
•
Temperature
•
Humidity
Max. 85 % rel. - non-condensing
•
Altitude
Max. 2500 mNN
Connecting load
The device can be used with the following electrical
connection values:
•
230 V ~/50 Hz 13 A single phase
•
110 V ~/60 Hz 16 A single phase
DANGER
When choosing the installation site
take into consideration that for
servicing a suitable limitation of the
laser area is possible (see the regula-
tions about accident prevention for
laser radiation BGV B2 or equivalent
national or international regulations
(e.g. EC Directive 60825 or IEC Publi-
cation 825).
Operation
Ambient temperature
- Full load: max. 35°C
- Partail load: max. 40°C
Storage
If the device contains cooling
water do not store or transport it
below 3°C (frost risk).
CAUTION
Never add antifreeze solution to
deionised cooling water!
NOTE
The device is manufactured according to DIN EN
61010, protection class 1 and is only to be
connected to grounded power sources (power
sockets with protective contacts).