Function test of the EMERGENCY STOP button
1.
Make sure that the laser has not been opera-
ted for a few minutes before starting the test
(or only at a very low output). Otherwise, there
is the danger that the laser lamp may overheat.
2.
Press the EMERGENCY STOP button to the
left of the stereo microscope.
Only the laser power supply is turned off,
the water pump is still running.
Setting / managing laser parameters
Fig. 27
Main menu
Use the [ - ] und [ + ] buttons to change the laser
parameters step by step (in the last indicated digit)
within the permitted limits.
The set parameters are used for the following
welding processes and can be saved for later uses
(
→
SAVING LASER PARAMETERS).
Fig. 28
Exceedance of parameter values
The current (new set) parameter values are
indicated between the buttons. Exceeded
parameter values are indicated in red characters in
the following cases:
•
permanently red, if a permitted limit for the
parameter has been reached or exceeded;
•
temporarily red if the laser is short term inoper-
able due to a change of the respective para-
meter.
Fig. 29
Opening the input window
1.
Touch any of the laser parameter values to
open the
→
ALPHANUMERIC INPUT WIN-
DOW for the input of a new value for the
respective parameter.