Status / error messages, causes and remedy
No. Symbol
Type
Meaning/Explanation
Measures
1
Working chamber door not or not
completely closed.
Completely close the working chamber
door.
If the symbol keeps on being indicated, ple-
ase contact the service department.
2
The Lamp is not yet ignited.
If the system is switched off and on
again very quickly, the laser's
power supply unit is discharged for
safety reasons. This can take seve-
ral minutes.
Wait until the capacitor bank is charged.
3
The capacitor bank is being
discharged.
Wait until the voltage of the capacitor bank
reaches the new value.
4
If the working chamber door is ope-
ned and then closed again while
the pedal switch is depressed, the
shutter closes for safety reasons.
Take your foot off the pedal switch.
5
IL
Water level in the tank too low.
Check water level.
Fill the tank to the maximum mark with DI
water.
If the symbol keeps on being indicated, ple-
ase contact the service department.
6
The water flow is too low.
Check if there is a kink in the cooling hose.
Check the circulating pump and permeabi-
lity of the filter by disconnecting the return
line from the filter and feeding it directly into
the supply tank.
Clean the coolant circuit (e. g. after a very
long period out of use or after a lamp bre-
akage).
If the symbol keeps on being indicated, ple-
ase contact the service department.
7
IL
The water flow is insufficient.
Check if there is a kink in the cooling hose.
Check circulation pump.
Clean the coolant circuit (e. g. after a very
long period out of use or after a lamp bre-
akage).
If the symbol keeps on being indicated, ple-
ase contact the service department.
8
IL
Maximum permissible temperature
of the cooling water exceeded.
Check the ventilators.
The ambient temperature is too high.
If the symbol keeps on being indicated, ple-
ase contact the service department.