32
860
76
D
B
-en/
03
35
5.0
00
32
English
⋅
LaserStar T Plus
en
As soon as the symbol
appears on
the display, the laser is ready for operation.
6.
Check the operation of the safety installations
(
→
section CHECK SAFETY EQUIPMENT)
Check safety equipment
Safety information
Visual inspection of the laser protective
window
The laser protective window should not have any
scratches, fractures, chips, cracks or discoloration.
If the protective laser window becomes defective, it
needs to be covered and immediately exchanged
with an intact window.
Make sure that the laser protective window is firmly
seated in the frame.
The laser protective window may only be cleaned
with clean water with added commercially available
dishwashing cleaning agent. Never use alcohol,
gas, acetone or similar solvents or hard or sharp-
edged objects.
Function test of the safety switch of the
working chamber door
Procedure
1.
Make sure that the working chamber door is
completely closed and the Laser key is in 1
position (middle position).
The shutter status display
(on the touch screen in the main menu) is
green; the shutter is not released.
2.
Touch the shutter status display.
The shutter status display switches to red
;the shutter is released, and
all safety circuits are closed.
3.
Open the working chamber door.
The shutter is closed immediately and the
shutter status display changes to
when the working chamber
door is open even slightly.
If you open the working chamber door too slow (so
that the debounce time is exceeded) an interlock is
triggered and the shutter status display switches to
.
NOTE
Repeatedly turning the unit on and off causes a
temperature-related delay of the laser firing since
the energy saved in the laser must be reduced
each time via resistors.
This delay can last up to several minutes.
Wait for at least two minutes before turning the
unit back on.
DANGER
Burning of eyes or skin
Check the condition of the laser
protective glass and the function of
the safety switches of the working
chamber door before the start of
each shift.
If the laser protective glass is
damaged (see following section) or if
the above mentioned safety switch
does not function properly, there is a
danger that a hazardous laser
radiation will exit the working
chamber.
In the cited cases, immediately turn
off the device and notify the respon-
sible service department.
If you suspect that you have been
burned, especially your eyes,
immediately notify the laser safety
supervisor or the safety expert and
consult a doctor if necessary.
DANGER
Burning of eyes or skin!
If the shutter is not closed as
described above, uncontrolled laser
radiation may be emitted.
In this case, immediately shut off the
device and notify the responsible
service.