Note
W 005
Internal battery empty
Notify Customer Service.
W 010
Chamber leak
Clean seals (
→
page 74).
W 011
Pressure increase too slow
Check compressed air supply
≥
5 bar (100 l/min).
Install compressed air reservoir, as needed
(
→
page 16).
Clean seals (
→
page 74).
W 012
Pressure increase too fast
* Notify Customer Service.
W 013
Maximum vacuum was not
reached
Check compressed air supply
≥
5 bar (100 l/min).
Clean the ejector nozzle (
→
p. 76).
Replace the air filter (
→
p. 77).
W 014
Maximum mould pressure
was not reached
Check compressed air supply
≥
5 bar (100 l/min).
Install compressed air reservoir, as needed
(
→
page 16).
Clean seals (
→
p. 74).
W 015
Critical cooling water tem-
perature
* Allow cooling water to cool down while the device is on.
W 016
Chamber not tight
Clean seals (
→
page 74). If the service message still
persists, notify Customer Service.
Error message
Cause
Remedy
Error message
Cause
Remedy
H 001
Initialization of the internal
USB flash drive
−
H 077 (every
8,000 castings)
Service required
Notify Customer Service.
H 099 (after
10,000 castings)
Service required
Notify Customer Service.
Summary of Contents for Nautilus CC plus
Page 7: ...English Nautilus CC plus Nautilus T 7 86105 BA en 00 en Unpacking Packing REF 17909 ...
Page 42: ...42 86105 BA en 00 English Nautilus CC plus Nautilus T en ...
Page 43: ...Nautilus CC plus en English ...
Page 59: ...Nautilus T en English ...
Page 83: ...English Nautilus CC plus Nautilus T 83 86105 BA en 00 en NAUTILUS T ...
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......