12
Non-Contact Voltage Sensing
1. To avoid personal injury or damage to the meter, do not test on un-
insulated wires.
2. Buzzer will sound and screen will display “OL” when detecting ac Voltage
above 90V ac.
3. Do not test on hazardous live wires higher than AC 600V
4. Before and after hazardous voltage measurements, test the Meter by
approaching to a known source such as a line AC voltage or outlet to
determine proper operation. See below figure.
5. At NCV mode, no test lead connections are required for NCV
measurement.
Measuring distance
8mm
Buzzer will sound when the detected voltage is
≥
90V, and the buzzer will be
on.The distances between the wire and the meter should be
≤
8mm.
X�
The voltage indication in NCV mode is not sufficient to assure safety.
This function is not suitable to test absence of voltage.