16
[1] 8 to 10 A up to 20 minutes, 10 minutes off minimum.
[2] >10 to 20 A overload for 30 seconds maximum, 10 minutes off minimum
[3] >10 to 20 A for 30 seconds accuracy ±(2.0%+3LSD).
8. AC Current Measurement
Range
Resolution
Accuracy
μA
400.0μA
[4]
0.1μA
±(1.2%+3LSD)
4000μA
[4]
1μA
mA
40.00mA
[4]
10μA
400.0mA
[4]
0.1mA
A
4.000A
[4]
1mA
±(1.5%+3LSD)
20.00A
[1, 2, 3, 4]
10mA
Function
Overload Protection
Overload
mA /μA
F1 fuse, 0.5A H 1000V fast-fuse,
(6.3×32)mm
N/A
A
F2 fuse, 11A H 1000V fast-fuse,
(10×38)mm
20 A overload for 30 seconds
maximum, 10 minutes off
minimum
[1] 8 to 10 A up to 20 minutes, 10 minutes off minimum.
[2] >10 to 20 A overload for 30 seconds maximum, 10 minutes off minimum
[3] > 10 to 20 A for 30 seconds accuracy ±(4%+5LSD).
[4] Frequency response: 45Hz ~ 400Hz
9. Temperature Measurement
Range
Resolution
Accuracy
-40 – 0°C
0.1°C @ < 400°C
1°C @
≥
400°C
±(10%+4 LSD)
>0 – 100°C
±(1.2%+3 LSD)
>100 – 1000°C
±(2.5%+2 LSD)
-40 – 32°F
0.1°F @ < 752°F
1°F @
≥
752°F
±(20%+6 LSD)
>32 – 212°F
±(1.8%+6 LSD)
>212 – 1832°F
±(2.5%+4 LSD)
Overload protection:
600V
K type (nickel-chromium/nichrosi) thermocouple must be used for temperature
measurement.
Summary of Contents for 0095969800341
Page 2: ......
Page 25: ...19 F1 9V ba tte ry F2 ...
Page 26: ......
Page 50: ......
Page 74: ......
Page 98: ......
Page 121: ...19 F1 9V ba tte ry F2 ...
Page 122: ......
Page 146: ......
Page 169: ...19 F1 9V ba tte ry F2 ...
Page 170: ......
Page 193: ...19 F1 9V ba tte ry F2 ...
Page 194: ......
Page 217: ...19 F1 9V ba tte ry F2 ...
Page 218: ......
Page 241: ...19 F1 9V ba tte ry F2 ...
Page 242: ...20 ...
Page 266: ......
Page 289: ...19 10 A indgangsterminal F2 sikring 11 A H 1000 V hurtig sikring 10 38 mm F1 9V ba tte ry F2 ...
Page 290: ......