C-4
Delay is necessary after issuing a setup command, due to theoutput voltage slew of 100v per 500ms.
On dual range Power Supplies, this command selects the LO or HI voltage range. Dual Ranges are 0-135 volts and
0-135/0-270 volt. If this command is omitted from the setup string the unit will default to the lowest available range.
On single range units this command will be ignored.
NOTES:
1. The setup command specifies the voltage and frequency of the AC power Supply output. The AC Power Supply
responds only to the last setup command entry. Its memory does not retain previously entered setup commands.
2. During EIA-232 remote operation, the unit's front panel VOLTS and FREQ must be fully CCW. If the controls are
not set fully CCW, then the output voltage and frequency will be a sum of the programmed value and the value set
with the front controls. Front Panel OUTPUT switch must be set to OFF position.
Front Panel RANGE switch must be set to LO.
CILL SOFTWARE COMMANDS
CILL COMMANDS
EXPLANATION
FETCH DATA COMMAND:
FTH VOLT <cr><lf><eos>
The AC supply responds to this command by transmitting the RMS voltage
(in volts) measured at its output, in decimal format as follows:
<sp><digit><digit><digit><dp><digit><digit><cr><lf><eos>
FTH CURR <cr><lf><eos>
The AC Supply responds by transmitting the current (in amps) measured at
its output, in decimal format, as follows:
<sp><digit><digit><dp><digit><cr><lf><eos>
FTH FREQ <cr><lf><eos>
The AC Supply response by transmitting the frequency (in hertz) measured
at its output, in decimal format, as follows:
<sp><digit><digit><digit><cr><lf><eos>
CLS :CH0 <cr><lf><eos>
Closes output relay contacts. Connects AC Power Supply outputs to output
terminal BLock. The setup command specifiesthe voltage and frequency.
The setupcommand must be entered before the CLS :CH0 command.
OPN :CH0 <cr><lf><eos>
Opens output relay contacts. Disconnects output of AC Power Supply from
output connector. Ac Power Supply retains voltage and frequency assigned
by setup command.
RST ACS:CH0
Resets AC Power Supply to quiescent. The output relay opens and error
messages are erased.
Catastrophic error messages are cleared by reading the Power Supply's
response, by use of the Status command.
CNF <cr><lf><eos>
Instructs AC Power Supply to perform internal confidence test. STA
command transmits the message. The AC Power Supply responds with one
of the following:
Pass
<sp><cr><lf><eos>
CILL SOFTWARE COMMANDS (cont)
Summary of Contents for BL3300 Series
Page 9: ...3 BL3300 SERIES MANUAL 01 28 2013 ALL MANUAL ADDENDUMS WILL FOLLOW THIS PAGE ...
Page 15: ...9 BL3300 SERIES MANUAL 01 28 2013 SECTION 2 UNPACKING AND INSTALLATION ...
Page 17: ...11 BL3300 SERIES MANUAL 01 28 2013 FIGURE 2 2 ...
Page 20: ...14 BL3300 SERIES MANUAL 01 28 2013 ...
Page 33: ...3 PHASE POWER CHASSIS MECHNICAL OUTLINE ...
Page 34: ...BL3300 CONTROL CHASSIS MECHANICAL OUTLINE ...
Page 35: ...BL33XX SERIES SINGLE PHASE POWER CHASSIS OUTLINE DRAWING 19 0 TYP 17 TYP 7 0 MAX ...
Page 54: ...6 1 SECTION 6 PARTS LIST AND DRAWINGS f ...
Page 56: ......
Page 57: ......
Page 58: ......
Page 60: ......
Page 62: ......