background image

is a 

BEK

 brand.

4.

 The nebulizer can be connected to the compressor by using the air hose with connector. 

5.

 Plug the DC adapter (optional) into the wall outlet; plug the power cord of the adapter into the socket on the 

side of the compressor.

6.

 Press the power switch to turn on the device. Begin treatment according to your physician’s instructions. 

7.

 Inhale and exhale naturally. Be sure to stop when the medication is exhausted.

Cleaning and Maintenance

Symptom

Probable Cause

Solution

The device does not work
or the device is on but 
nebulizes weakly. 

The device is on but does not 
nebulize.   

The plug is improperly inserted into the 
device.

Compressor switch is off.

Air hose is bent.

Turn the switch on.

Remove any bend or kink in the air hose.

The spray nozzle does not place in its 
position. 

Place the spray nozzle in its correct position.

No medicine is left.

The medicine container and cap does not 
assemble well.  

The texture of medicine may be thick.

Add the appropriate amount of medicine 
prescribed by your physician to the medicine 
container. 

Re-assemble the medicine container and cap 
and twist them well.

Slightly shake the medicine well and then 
nebulization will be produced. 

Re-plug again and make sure that the plug is 
properly inserted into the device. 

Troubleshooting

Specifications

Symbol Information

Note

After the treatment, perform the followings in order: 
1. Turn off the device.
2. Dis-connect the air hose.
3. Unplug the adapter.
4. Remove other used accessories.
5. Proceed with the cleaning procedure as directed. 

Instructions on Mouthpiece and Mask
Mouthpiece
1.

 Attach the mouthpiece to the top of the cap.

2.

 Close your lips around the mouthpiece and start treatment as recommended by your doctor.

Mask
1.

 Insert the mask into the cap.

2.

 Place the mask onto your face and start treatment as recommended by your doctor.

Before Cleaning

Make sure that the nebulizer is turned off before disassembly.

For Accessories
1.

 After you have disconnected the air hose, remove the mouthpiece / mask, cap, spray nozzle and 

medicine container.

2.

 Discard any remaining medications.

3.

 Wash mouthpiece / mask, cap, spray nozzle and medicine container with running water or soak in 

warm water for 15 minutes.

4.

 Gently wipe off excess water with a clean dry cloth and allow to air dry on a clean paper towel.

Clean the unit and accessories (adapter not included) as directed below:

ITEM

WHEN TO CLEAN (

)

Accessories

Compressor unit

 for the first time   

 after each use   

 a period of no use

○ 

for the first time   

 after each use   

 a period of no use

For the Compressor Unit

1. Use alcohol cotton swabs to wipe the outer casing of the compressor.

2. Let it air dry on a clean paper towel.

• The above-named accessories can be regarded as consumable items; please replace with new 
ones if they are dirty and obstructed. To prevent cross-contaminations, do not share the 
accessories with others.
• Use only the aerosol nebulizer kit manufactured by Lifechek for appropriate assembly.  
   Do not use accessories which are not supplied or recommended
   Other cables and accessories may negatively affect EMC performance.

Note

Storage

• Always handle your nebulizer with care.
• Disconnect the power plug from the wall outlet when not in use.
• Keep your nebulizer out of children’s reach.
• If you store your nebulizer, do not bend the air hose.
• If you store your nebulizer, try to keep it in the following environmental ranges:
a) Temperature: -25 °C to 70 °C (-13 °F to 158 °F), 
b) Relative humidity: 10% to 95% 
• If possible, stroe your nebulizer in a well-ventilated room.

Disposal

The used unit should be treated as contaminated that may carry a risk of infection during nebulization. Dispose of 
the device, components, used batteries, and optional accessories according to your local regulations.

Model No.:

                            TD-7013

Dimension & Weight:

           125 x 115 x 46 mm; 220 g

Power Source:

                      DC 12 V power adapter

 • 

Input:

                              100 to 240V, AC

 • 

Output:

                            12V, 1A, DC

IP Classification:

                  IP21

Mode of Operation:

              30 mins on/ 30 mins off  

Nebulization Rate:

                0.35 ml/min

Medication Capacity:

           6 ml

MMAD:

                                  Approximately 5 microns

Power Consumption:

           12 W

Compressor Pressure Range:

 97~124kPa (14~18 Psi)

Operating Pressure Range:

     35 to 55kPa (5 to 8 Psi)

Maximum Flow Rate:

              4 LPM (l/min)

Aerosol output:                        

 1.80 ± 0.02 (ml)*  

Aerosol output rate:

               0.21 ± 0.02 (ml/min)** 

*: Continue the treatment until the medication cup is empty or the mist stops.
**: The treatment time for aerosl output rate is 1 min.

Operating conditions:

             5°C to 40°C (40°F to 104°F), 15% to 93% relative humidity, 700 hPa to 1060 hPa

Storage conditions:

                -25°C to 70°C (-13°F to 158°F), 10% to 95% relative humidity.

Manufacturer

Type BF Equipment

Serial number

Ingress protection rating

Humidity limitation

Symbol

Referent

Consult instructions for use

Temperature limitation

Caution, consult accompanying documents

Authorized representative in the

European Community

Symbol

IP21

CE mark

Referent

RoHS Compliance

DISPOSAL: Do not dispose this product as 

unsorted municipal waste. 

Collection of such waste separately for special 

treatment is necessary.

Summary of Contents for Lifechek TD-7013

Page 1: ...rın olup olmadığını kontrol ediniz Not Hijyenik sebeplerden dolayı lütfen ağızlığı sadece bir kișiye özel olacak șekilde kullanınız ve kullanıcı değiștirmeyiniz Sadece yukarıda listelenen parçaları kullanınız farklı bir parçanın kullanılması tehlike yaratabilir Kompresörün ömrü 3 yıldır Cihazın sık kullanımı ömrünü kısaltabilir 12 Watt sürekli çıkıș gücü sağlayan güç adaptörünün kullanılması öneri...

Page 2: ...iden birleștiriniz ve yavașça çevirerek kapatınız İlacı yavașça sallayıp karıștırdıktan sonra nebulizasyon gerçekleșecektir Kabloyu cihaza yeniden takınız ve iyice yerleștirdiğinizden emin olunuz Sorun Giderme Özellikler Sonuçların Kümülatif Dağılım Grafiği Model No TD 7013 Boyut Ağırlık 125 mm x 115 mm 220 g Güç kaynağı DC 12V Giriș 100 ila 240V AC Çıkıș 12V 1A DC IP sınıflandırması IP21 Çalıșma ...

Page 3: ...ing or swallowing The accessible materials used in the device will not cause the potential allergic reactions to skin Do not try to modify the device to prevent any dangers Do not place the device in liquid nor put it where it could fall into liquid If the device becomes wet unplug the device before touching it Do not use the device if it is not working properly or if it has suffered any damage Do...

Page 4: ...O CLEAN Accessories Compressor unit for the first time after each use a period of no use for the first time after each use a period of no use For the Compressor Unit 1 Use alcohol cotton swabs to wipe the outer casing of the compressor 2 Let it air dry on a clean paper towel The above named accessories can be regarded as consumable items please replace with new ones if they are dirty and obstructe...

Page 5: ...level 3 Vrms 0 15 MHz 80 MHz 6 Vrms in ISM and amateur radio bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Compliance level Electromagnetic environment guidance for home and professional healthcare environment Recommended separation distance between portable and mobile RF communications equipment and the Lifechek TD 7013 The Lifechek TD 7013 is intended for ...

Page 6: ...rli olmayabilir Elektromanyetik yayılmalar yapılardan nesnelerden ve insanlardan yapılan emilim ve yansımalardan etkilenir İletilen RF IEC 61000 4 6 Yayılan RF IEC 61000 4 3 Tașınabilir ve mobil RF iletișim ekipmanı TD 7013 ün kablolar dahil herhangi bir bölümüne vericinin frekansına uygulanabilir denklemden hesaplanan önerilen ayırma mesafesinden daha yakın kullanılmamalıdır Önerilen ayırma mesaf...

Reviews: