Plot of cumulative size distribution of results 

0.1

1

10

100

0

20

40

60

80

100

0

20

40

60

Cumulative Fraction (%)

Mean Mass Fraction (%)

AeroDynamic Diameter (µm)

individual tests

mean

Reference to Standards:

• Electric Safety Standards EN 60601-1
• Electromagnetic Compatibility according to EN 60601-1-2
• Particle Size according to EN 13544-1, clause 6.8.2 
The device is a class IIa medical device according to Medical Device Directive 93/42/EEC.

• Please contact your dealer for assistance with any other difficulties.
• Performance may vary with medication such as suspensions or high viscosity. See    
medication supplier’s data sheet for further details.
• Performance information provided by the manufacturer in accordance with the  

 

standard EN 13544-1:2007 may not apply to drugs supplied in suspension or high  

 

viscosity form. Please consult the drug supplier for more information.

Note

NOTE  U

T

 is the a.c. mains voltage prior to application of the test level.

Manufacturer’s declaration-electromagnetic immunity

The TD-7013 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.
The customer or the user of the TD-7013 should assure that it is used in such an environment.

Immunity

test

Electrostatic discharge
(ESD) IEC 61000-4-2

Electrical fast 
transient/burst IEC 
61000-4-4

Surge IEC 61000-4-5

Voltage Dips, short 
interruptions and voltage 
variations on power 
supply input lines IEC 
61000-4-11

Power frequency (50, 60 
Hz) magnetic field IEC 
61000-4-8

Contact: ±8 kV 
Air±2 kV,±4 kV,±8 kV,±15 
kV
+ 2kV for power supply 
lines
+ 1kV for input/output 
lines
+ 0.5kV, +1kV line(s) to 
line(s)
+ 0.5kV, +1kV,+ 2kV 
line(s) to earth
Voltage dips:
0 % U

T

; 0,5 cycle             

0 % U

T

; 1 cycle

70 % U

T

; 25/30 cycles

Voltage interruptions:
0 % U

T

; 250/300 cycle

30 A/m
50 Hz or 60 Hz

Contact: ±8 kV 
Air±2 kV,±4 kV,±8 
kV,±15 kV
+ 2kV for power supply 
lines
Not applicable

+ 0.5kV, +1kV line(s) to 
line(s) 
Not applicable

Voltage dips:
0 % U

T

; 0,5 cycle             

0 % U

T

; 1 cycle

70 % U

T

; 25/30 cycles

Voltage interruptions:
0 % U

T

; 250/300 cycle

30 A/m
50 Hz

Floors should be wood, concrete or ceramic tile. If floors are 
covered with synthetic material, the relative humidity should 
be at least 30%.
Mains power quality should be that of a typical home 
healthcare environment.

Mains power quality should be that of a typical home 
healthcare environment.

Mains power quality should be that of a typical home 
healthcare environment. If the user of the TD-7013 requires 
continued operation during power mains interruptions, it is 
recommended that the TD-7013 be powered from an 
uninterruptible power supply or a battery.

The TD-7013 power frequency magnetic fields should be at 
levels characteristic of a typical location in a typical home 
healthcare environment.

IEC 60601 

test level

Electromagnetic environment-guidance

(for home and professional healthcare environment)

Compliance

level

NOTE 1

  At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2

  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, 

objects and people.

Manufacturer’s declaration-electromagnetic immunity

The TD-7013 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.
The customer or the user of the TD-7013 should assure that it is used in such and environment.

Conducted RF 
IEC 61000-4-6

Radiated RF IEC 
61000-4-3

Portable and mobile RF communications 
equipment should be used no closer to any part of the 
TD-7013 including cables, than the recommended separation 
distance calculated from the equation applicable to the 
frequency of the transmitter.

Recommended separation distance:
d = 1,2 √P
d = 1,2 √P  80MHz to 800 MHz
d = 2,3 √P  800MHz to 2,7 GHz 

Where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation distance 
in metres (m).

Interference may occur in the vicinity of equipment marked 
with the following symbol:

Immunity

test

3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: 
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz 

80 % AM at 1 kHz 

10 V/m
80 MHz – 2,7 GHz 
80 % AM at 1 kHz

IEC 60601 

test level

3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: 
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz 

80 % AM at 1 kHz 

10 V/m
80 MHz – 2,7 GHz 
80 % AM at 1 kHz 

Compliance

level

Electromagnetic environment-guidance

(for home and professional healthcare environment)

Recommended separation distance between

portable and mobile RF communications equipment and the Lifechek TD-7013

The Lifechek TD-7013 is intended for use in an electromagnetic environment (for home and professional healthcare) in which radiated RF 
disturbances are controlled. The customer or the user of the Lifechek TD-7013 can help prevent electromagnetic interference by maintaining a 
minimum distance between portable and mobile RF communications equipment (transmitters) and the Lifechek TD-7013 as recommended 
below, according to the maximum output power of the communications equipment.

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated 
using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) 
according to the transmitter manufacturer.

NOTE 1

  At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2

  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, 

objects and people.

Separation distance according to frequency of transmitter

m

Rated maximum output

power of transmitter

(W)

0,01

0,1

1

10

100

150 kHz to 80 MHz

d =1,2√P

0,12
0,38

1,2
3,8

12

80 MHz to 800 MHz

d =1,2√P

0,12
0,38

1,2
3,8

12

800 MHz to 2,7 GHz

d =2,3√P

0,23
0,73

2,3
7,3

23

The TD-7013 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.
The customer or the user of the TD-7013 should assure that it is used in such an environment.

Emission test

Compliance

Electromagnetic environment-guidance 

(for home and professional healthcare environment)

RF emissions CISPR 11

RF emissions CISPR 11
Harmonic emissions 
IEC 61000-3-2
Voltage fluctuations / flicker emissions 
IEC 61000-3-3

Group 1

Class B
Class A

Compliance

The TD-7013 uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely to 
cause any interference in nearby electronic equipment.
The TD-7013 is suitable for use in all establishments, 
including domestic establishments and those directly 
connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes.

Manufacturer’s declaration-electromagnetic emissions

Manufacturer’s declaration-electromagnetic immunity

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment

385

450

710
745
780

810
870
930

1 720
1 845
1 970

2 450

5 240
5 500
5 785

NOTE 

 If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be 

reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.

a)

  For some services, only the uplink frequencies are included.

b)

  The carrier shall be modulated using a 50 % duty cycle square wave signal.

c)

  As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would

be worst case.

The Lifechek TD-7013 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.
The customer or the user of the 

Lifechek TD-7013

 should assure that it is used in such an environment.

Test

frequency

(MHz)

Band 

a)

(MHz)

Service 

a)

Modulation 

b)

Compliance

LEVEL

(V/m)

(for home

healthcare)

27

28

9

28

28

28

9

Immunity

test

LEVEL

(V/m)

27

28

9

28

28

28

9

Distance

(m)

0,3

0,3

0,3

0,3

0,3

0,3

0,3

Maximum

power

(W)

1,8

2

0,2

2

2

2

0,2

Pulse modulation 

b)

18 Hz

FM 

c)

±5 kHz deviation
1 kHz sine

Pulse modulation 

b)

217 Hz

Pulse modulation 

b)

18 Hz

Pulse modulation 

b)

217 Hz

Pulse modulation 

b)

217 Hz

Pulse modulation 

b)

217 Hz

TETRA 400

GMRS 460, FRS 460

LTE Band 13, 17

GSM 800/900, TETRA 800, 
iDEN 820, CDMA 850, 
LTE Band 5

GSM 1800; CDMA 1900;
GSM 1900; DECT; LTE Band
1, 3, 4, 25; UMTS

Bluetooth, WLAN, 802.11 b/g/n, 
RFID 2450, LTE Band 7

WLAN 802.11
a/n

380 – 390

430 – 470

704 – 787

800 – 960

1 700 – 1 990

2 400 – 2 570

5 100 – 5 800

B.E.K.

 

Sağlık Ürünleri Eğitim Turizm Emlak ve Danıșmanlık Hizmetleri San. ve Tic. Ltd. Ști. 

Ahmet Yesevi Mah. Sanayi Cad. Akarsu İș Merkezi No: 523/21 Nilüfer/Bursa/TÜRKİYE 
Tel: +90 444 9 235   e-posta: info@beksaglik.com.tr   www.lifechek.com.tr

Summary of Contents for Lifechek TD-7013

Page 1: ...rın olup olmadığını kontrol ediniz Not Hijyenik sebeplerden dolayı lütfen ağızlığı sadece bir kișiye özel olacak șekilde kullanınız ve kullanıcı değiștirmeyiniz Sadece yukarıda listelenen parçaları kullanınız farklı bir parçanın kullanılması tehlike yaratabilir Kompresörün ömrü 3 yıldır Cihazın sık kullanımı ömrünü kısaltabilir 12 Watt sürekli çıkıș gücü sağlayan güç adaptörünün kullanılması öneri...

Page 2: ...iden birleștiriniz ve yavașça çevirerek kapatınız İlacı yavașça sallayıp karıștırdıktan sonra nebulizasyon gerçekleșecektir Kabloyu cihaza yeniden takınız ve iyice yerleștirdiğinizden emin olunuz Sorun Giderme Özellikler Sonuçların Kümülatif Dağılım Grafiği Model No TD 7013 Boyut Ağırlık 125 mm x 115 mm 220 g Güç kaynağı DC 12V Giriș 100 ila 240V AC Çıkıș 12V 1A DC IP sınıflandırması IP21 Çalıșma ...

Page 3: ...ing or swallowing The accessible materials used in the device will not cause the potential allergic reactions to skin Do not try to modify the device to prevent any dangers Do not place the device in liquid nor put it where it could fall into liquid If the device becomes wet unplug the device before touching it Do not use the device if it is not working properly or if it has suffered any damage Do...

Page 4: ...O CLEAN Accessories Compressor unit for the first time after each use a period of no use for the first time after each use a period of no use For the Compressor Unit 1 Use alcohol cotton swabs to wipe the outer casing of the compressor 2 Let it air dry on a clean paper towel The above named accessories can be regarded as consumable items please replace with new ones if they are dirty and obstructe...

Page 5: ...level 3 Vrms 0 15 MHz 80 MHz 6 Vrms in ISM and amateur radio bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Compliance level Electromagnetic environment guidance for home and professional healthcare environment Recommended separation distance between portable and mobile RF communications equipment and the Lifechek TD 7013 The Lifechek TD 7013 is intended for ...

Page 6: ...rli olmayabilir Elektromanyetik yayılmalar yapılardan nesnelerden ve insanlardan yapılan emilim ve yansımalardan etkilenir İletilen RF IEC 61000 4 6 Yayılan RF IEC 61000 4 3 Tașınabilir ve mobil RF iletișim ekipmanı TD 7013 ün kablolar dahil herhangi bir bölümüne vericinin frekansına uygulanabilir denklemden hesaplanan önerilen ayırma mesafesinden daha yakın kullanılmamalıdır Önerilen ayırma mesaf...

Reviews: