Check the provided water temperature:
The mixers are supplied with a water temperature calibration setting of + 41° C supplied, with a hot
water supply temperature of 60-65 ° C, and a cold water temperature of 10-15 ° C. If the water is
too hot or too cold after installation a readjustment of the temperature controller is required. This is
done as follows:
Open the valve for bath filling and set the desired temperature. (A press of the security lock may
be required).
Connect the bath filling to the mixer. (The main section is still under pressure).
Remove the middle screw on the knob. Pull the knob straight out without you adjusting the
spindle setting.
Bring the knob with the internal fitting to the right side of the guide ring.
Now attach the screw to the knob.
Check the provided water temperature.