Prism Medical Ltd. V1-2015
28
CAUTION:
NOTICE:
OK
1. The container with the disinfectant can be
accessed via the inspection cover on the back
side of the AVERO Premium Plus lifting tub.
When refilling and replacing the container
make sure that the hose reaches the bottom of
the container.
2. Take the disinfection handheld showerhead
(yellow hose) and hold it so that the disinfecting
solution sprays into the tub.
Press
key 1
to start the disinfection. Use care
when spraying the inside of the tub. Let the
disinfectant set for long enough.
(see manufacturer recommendations).
Then rinse the tub thoroughly with clean water
(with the patient handheld showerhead).
3. Regularly check
(min. once a week) if there is still enough
disinfectant in the container. If necessary, change
the container. Make sure that the hose reaches
the bottom of the container.
Removable
inspection cover for
refilling the
disinfectant
(removable)
Handheld
showerhead
disinfection
Only use disinfectant after the patient has left the tub.
Follow the manufacturer instructions strictly for the
disinfectant used. Avoid direct contact with the
concentrated product. Where appropriate, protect the
skin and eyes with gloves and goggles.
4. Changing the Disinfection Container
When changing the disinfectant container open the
inspection panel with the key provided. Remove
the empty disinfectant container. Pull the suction
hose out of the empty disinfectant container. Plug
the hose into the new disinfectant container. Ensure
the suction hose reaches the bottom of the
container. Place the disinfectant container in the
holder. Close the inspection panel. Check the
functioning of the disinfectant equipment
.
To ensure that no water remains in the nozzles of the
air spa, the fan will continue to run for about 30
seconds.
DO NOT interrupt this "dry cycle" Wait until the fan
shuts off automatically!
Summary of Contents for 810142501
Page 1: ...1 Plumbing CSA B45 5 11 IAPMO 7124 2011 Electrical CSA C22 2 No 60601 1 ...
Page 6: ...Prism Medical Ltd V1 2015 6 2 Introduction WARNING ...
Page 31: ...Prism Medical Ltd V1 2015 31 11 Troubleshooting and customer service Problem Solution ...
Page 35: ...Prism Medical Ltd V1 2015 35 ...
Page 36: ...Prism Medical Ltd V1 2015 36 ...
Page 37: ...Prism Medical Ltd V1 2015 37 ...