The disinfection key turns the disinfection device on or off.
To activate disinfection, press this key.
To turn off the disinfection you press the key again.
The disinfection device can be turned on only when the tub is emptied and all other
functions are turned off.
This applies to, for example, the fan, dry-out light, handheld showerhead and filling features.
A disinfection procedure must last for at least 15 seconds, otherwise the system starts from
a faulty operation.
The power LED (14) shows the activation of disinfection through slow flashing.
The height adjustment feature of the tub is not affected by the lock and is still in function.
After disinfection, the exposure time of 5 minutes and the subsequent release of all
functions begin.
On request, the exposure time can be optionally converted to 8 or 10 minutes by a service
technician.
To cancel the exposure time (locking of other functions), press the disinfection key for 5
seconds.
On the built-in digital display the following is shown:
Display (2) = Handheld showerhead temperature
Display (3) = Filling temperature
Display (4) = Bath water temperature
The display is shown in 0.5 °C increments.