Installation
Operating manual
KINET
V1.4 E
Page 12 of 40
6 Installation
The bath tub is supplied ready for use. Use the level and the wrench to vertically align
the bath tub. The tub legs can be screwed in or out to adjust their height. Please make
sure that all legs are in contact with the floor and that the bath tub stands stable.
Prior to the assembly of the bath, the local connection pipes must be sufficiently
flushed before the tilting bath tub is connected. To avoid soiling of the pipes and
fittings, dirt filters must be installed on site.
The bath tub must be connected through 1/2” hoses. Please observe the attached
connection diagram. The bath tub must be connected by a qualified and competent
professional.
CAUTION
To connect the footbath to the water supply, dirt filters must be used
on site. Please contact your plumber to this effect.
The image shows the water connections
of the
KINET
tilting bath tub.
6.1 Electrical connection
The connection of the bath tub must be executed in accordance with the technical
drawings issued by BEKA Hospitec either the guidelines for the set-up and layout of
wet areas and the current VDE standards. All supply and drain hoses as well as the
electrical cables must be installed so as to avoid the danger of tripping.
The
KINET
is connected to the power supply network through a shock-proof
protective socket featuring the legally prescribed splash guard (IP class), 230 V
AC/50Hz, 3 x 1.5 mm², 1.5 KW, 16 Ampere, fault-current circuit breaker 30 mA,
equipotential bonding (in Germany) to DIN VDE 0107.
The bath tub must be
connected by a qualified and competent professional.
6.2 Protective conductor
When protection class I equipment is used, the quality of the protective conductor of
the installation is highly important. Please note that in this respect regulations and
prescriptions have been specified in many countries by the national authorities.
WARNING
In order to avoid the risk of electroshock, this device only must be
connected to a power supply network with protective conductor.
= cold water
= hot water