Appendix
Operating manual
KINET
V1.4 E
Page 34 of 40
Guidance and manufacturer’s declaration - electromagnetic immunity (Table 202)
The
KINET
been designed for use in the hereafter listed ELECTROMAGNETIC ENVIRONMENTS. The
customer or the user of the
KINET
must ensure that the appliance is used in such environment.
Immunity testing
IEC 60601- Test level
Compliance level
Electromagnetic
environment guidance
Discharging of static
electricity (ESD) to IEC
61000-4-2
± 6kV contact discharge
± 8kV air discharge
± 6kV contact discharge
± 8kV air discharge
The floor must be in
wood, concrete or
ceramic tiles. In case of
floors in synthetic
material, the relevant air
humidity must be at
least 30%.
Rapid transient
interference
pulses/burst
IEC 61000-4-4
± 2 kV for power supply
cables
± 1 kV for input/output
cables
± 2 kV for power supply
cables
not applicable to input/
output cables
The quality of the
supply voltage should
match that of a typical
business or hospital
environment.
Overvoltage
IEC 61000-4-5
±/1 kV cable against cable
±/2 kV cable against
ground connection
±/1 kV cable against
cable
±/2 kV cable against
ground connection
The quality of the
supply voltage should
match that of a typical
business or hospital
environment.
Voltage drops, short
interruptions and
voltage fluctuations in
the power supply input
cables
IEC 61000-4-11
<5 % U
T
(>95 % drop of U
T
)
for 0.5 period
<40 % U
T
(>60 % drop of U
T
)
for 5 periods
<70 % U
T
(>30 % drop of U
T
)
for 25 periods
<5 % U
T
(>95 % drop of U
T
)
for 5 s
<5 % U
T
(>95 % drop of U
T
)
for 0.5 period
<40 % U
T
(>60 % drop of U
T
)
for 5 periods
<70 % U
T
(>30 % drop of U
T
)
for 25 periods
<5 % U
T
(>95 % drop of U
T
)
for 5s
The quality of the
supply voltage should
match that of a typical
business or hospital
environment.
Current frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
business or hospital
environment.
CAUTION U
T
is the mains AC voltage before the application of the test level.