9.2 Disinfecting the lift
You must carefully disinfect and rinse your product after each use to avoid the risk of
transmission and infection. For the manual disinfection of the surface, an isopropyl
alcohol solution or a customary disinfection aerosol (spray) can be used.
9.3 Sterilising the lift
The lift CARLO is
not
suitable for sterilization.
10 Checks/tests
In order to ensure a safe use of our product and the protection of the users and the
patients, BEKA Hospitec prescribes an annual safety check.
The execution of the safety checks and maintenance must be documented and proven
on request. Please use your inventory register to this effect.
We recommend a simultaneous maintenance of the device in order to conserve its full
value.
The checks may only be conducted by the manufacturer either by specialists
authorized by the manufacturer. Please contact the manufacturer upon expiration of
the validity date mentioned on the inspection sticker located on the rear side of the
device.
NOTE
In accordance with the UVV (accident prevention) regulations
of the German employer’s liability insurance association on
stationary equipment which is used in special locations or
installations, an annual check to the DGUV (German Statutory
Accident Insurance Association) Prescription 3 (BGV A3)
must be carried out on the product.
This check is only prescribed for Germany. In other countries, other requirements
might be applicable.
WARNING
Do not conduct any cleaning, maintenance or test activities
when the product is in use. This could cause danger to the
user and the patient.
Conduct a
daily
cleaning and disinfection of the product.
Conduct a
weekly
visual inspection of all components, the power cable, the hoses and
the connections. Conduct a functional test as well.
Conduct
every year
a maintenance, a safety check and a check to DGUV
prescription 3.