13.7.2 Front Castor Replacement
Image 1:
Required tools:
1 x 10 mm Allen
wrench
Image 2:
Additional parts/tools:
1 x front castor
(Order number
S9200350)
1 x thread lock fluid
(medium tight)
Image 3:
Put an object (e.g.
wooden block) under
the side where you
want to replace the
castor to free the
castor.
Image 4:
Unscrew the screw of
the castor using the
Allen wrench.
Image 5:
Remove the screw of
the castor. Hold the
castor back with your
hand!
Remove the castor
with the screw
Image 6:
Prior to screwing a
new castor, you must
apply thread lock fluid
(
medium tight at
least 21 Nm
) to the
thread of the screw.
Caution: the thread lock fluid must dry at least 3 hours before the device can be
loaded.
Should you use thread lock fluid from another manufacturer, the drying times
specified by this manufacturer must be observed!
Re-assemble in reversed order!